KIP-Veröffentlichungen

Jahr 2014
Autor(en) C. R. Bates, C. Pies, S. Kempf, L. Gastaldo, A. Fleischmann, C. Enss, S. Friedrich
Titel Development of MMC Gamma Detectors for Nuclear Analysis
KIP-Nummer HD-KIP 14-44
KIP-Gruppe(n) F4
Dokumentart Paper
Quelle J. Low Temp. Phys. 176, 631 (2014)
doi 10.1007/s10909-013-1063-7
Abstract (de)

Non-destructive assay (NDA) of nuclear materials would benefit from gamma detectors with improved energy resolution in cases where line overlap in current Ge detectors limitsNDAaccuracy.We are developing metallic magnetic calorimeter gamma-detectors for this purpose by electroplating ∼150μm thick Au absorbers into microfabricated molds on top of Au:Er sensors. Initial tests under non-optimized conditions show an energy resolution of ∼200eV FWHM at 60 keV. Monte Carlo simulations illustrate that this resolution is starting to be sufficient for direct detection of 242Pu in plutonium separated from spent nuclear fuel.