Jahr | 2016 |
Autor(en) | C. R. Bates, C. Pies, S. Kempf, D. Hengstler, A. Fleischmann, L. Gastaldo, C. Enss, S. Friedrich |
Titel | Reproducibility and calibration of MMC-based high-resolution gamma detectors |
KIP-Nummer | HD-KIP 16-55 |
KIP-Gruppe(n) | F4 |
Dokumentart | Paper |
Quelle | Appl. Phys. Lett. 109, 023513/1-4 (2016) |
doi | 10.1063/1.4958699 |
Abstract (en) | We describe a prototype γ-ray detector based on a metallic magnetic calorimeter with an energy resolution of 46 eV at 60 keV and a reproducible response function that follows a simple second- order polynomial. The simple detector calibration allows adding high-resolution spectra from differ- ent pixels and different cool-downs without loss in energy resolution to determine &gammay;-ray centroids with high accuracy. As an example of an application in nuclear safeguards enabled by such a &gammay;-ray detector, we discuss the non-destructive assay of 242Pu in a mixed-isotope Pu sample. |